V-2000
Method Updates The
V-2000 photometer makes use of internally stored calibration software,
which CHEMetrics periodically upgrades.
As calibrations are updated or new methods become available, a
simple upload procedure allows analysts to upgrade their V-2000
calibrations from their computers. The
step-by-step procedure is listed below and in Chapter 5 of the V-2000
Operator’s Manual. Current
Method Version: The
link below will upload the most current method version, Method Revision
No. 8.0. This method
revision became available on March 24, 2008 and reflected changes to the
following Vacu-vials methods:
To
determine which method revision is currently present on your V-2000
instrument, simply turn the instrument on.
As the instrument powers up, the method revision number is
displayed briefly. Equipment
You Will Need to Upgrade your V-2000: Connecting
a computer to the V-2000 requires the use of the RS-232 serial cable
supplied with the V-2000. This
cable has a special 3-pin connector on one end for the V-2000.
The RS-232 port is on the underside of the instrument.
When connecting the RS-232 cable to a computer, a serial
connector may be required (a 25 pin to 9 pin adapter is included with
the V-2000). If your
computer does not have a serial port, you will need to purchase a
separate adapter that will allow you to connect the V-2000 RS-232 serial
cable to the USB port on your computer.
The following USB to serial adapters are compatible with multiple
computer operating systems:
Procedure
for Method Revision Update:
It
is recommended that you print this page of instructions before
continuing. The
step-by-step procedure is also listed in the V-2000 Operator’s Manual,
Chapter 5 section 5-3 Uploading Method Revisions.
NOTE:
Uploading a new method revision does not overwrite any
user generated custom methods.
NOTE:
To abort the upload process, press the yes key again.
For further reference, see: V-2000 Photometer Operator’s Manual, Chapter 5 section 5-3 Uploading Method Revisions or contact Technical Support at 540-788-9026 / technical@chemetrics.com. Method
Update History:
|