V-2000 Method Updates

The V-2000 photometer makes use of internally stored calibration software, which CHEMetrics periodically upgrades.  As calibrations are updated or new methods become available, a simple upload procedure allows analysts to upgrade their V-2000 calibrations from their computers.  The step-by-step procedure is listed below and in Chapter 5 of the V-2000 Operator’s Manual.

Current Method Version:

The link below will upload the most current method version, Method Revision No. 8.0.  This method revision became available on March 24, 2008 and reflected changes to the following Vacu-vials methods:

Parameter

Kit Cat. No.

Program Code

Range

Reason for change

Hydrazine

K-5003

 89

0.10-1.20 ppm

Calibration update; accuracy improvement especially at low concentrations; reagent blank required

Oxygen 1

K-7503

140

0.20-2.00 ppm

Calibration update; accuracy improvement across test range

Oxygen 2

K-7513

141

2.0-15.0 ppm

Calibration update; accuracy improvement across test range

Phosphate 2

K-8513

159

0.30-8.00 ppm

Calibration update; PDL improvement

Sulfate

K-9203

174

8.0-100.0 ppm

Calibration update; accuracy improvement across test range

To determine which method revision is currently present on your V-2000 instrument, simply turn the instrument on.  As the instrument powers up, the method revision number is displayed briefly.

Equipment You Will Need to Upgrade your V-2000:

Connecting a computer to the V-2000 requires the use of the RS-232 serial cable supplied with the V-2000.  This cable has a special 3-pin connector on one end for the V-2000.  The RS-232 port is on the underside of the instrument.  When connecting the RS-232 cable to a computer, a serial connector may be required (a 25 pin to 9 pin adapter is included with the V-2000).  If your computer does not have a serial port, you will need to purchase a separate adapter that will allow you to connect the V-2000 RS-232 serial cable to the USB port on your computer.  The following USB to serial adapters are compatible with multiple computer operating systems:

Manufacturer

Manufacturer Part No.

Supplier

Supplier Part No.

Aten Technologies
aten-usa.com

UC232A

CDW Company
www.cdw.com

325871

CP Technologies
cptechusa.com

CP-US-03

CDW Company
www.cdw.com

767105

 

Procedure for Method Revision Update:

It is recommended that you print this page of instructions before continuing.

The step-by-step procedure is also listed in the V-2000 Operator’s Manual, Chapter 5 section 5-3 Uploading Method Revisions. 

NOTE:  Uploading a new method revision does not overwrite any user generated custom methods.

  1. With the V-2000 power off, connect the photometer to a computer using the supplied RS-232 serial cable.
  2. Press the power key to turn the meter on.
  3. Press the setup key and use the “up” or “down” keys until “UPLOAD” is displayed, then press the yes key.
  4. Meter should read “WAITING”.

NOTE:  To abort the upload process, press the yes key again.

  1. Initiate the download program: Click here to Download Revision No. 8.0 and follow the instructions provided in the file.
  2. When the program is complete, the instrument will briefly display “ALL DONE” along with the new method revision number.

For further reference, see: V-2000 Photometer Operator’s Manual, Chapter 5 section 5-3 Uploading Method Revisions or contact Technical Support at 540-788-9026 / technical@chemetrics.com.

Method Update History:

Method Revision No.

Date

Update

3.0

Jan 29, 2004

Original Version

4.0

Feb 24, 2004

Program 49 (HR COD): calibration update

5.0

Nov 12, 2004

Program 49 (HR COD): calibration update

6.0

May 23, 2005

Program 64 (DEHA 1): calibration update

7.0

Mar 10, 2006

Program 122 (Nitrate 4): calibration added

8.0

Mar 24, 2008

Programs 89 (Hydrazine), 140 (Oxygen 1), 141 (Oxygen 2), 159 (Phosphate 2), 174 (Sulfate): calibration updates